• DocumentCode
    3692274
  • Title

    Eliminating speckle noise with three-dimensional single-track-location shear wave elasticity imaging (STL-SWEI)

  • Author

    Peter Hollender;Samantha Lipman;Gregg Trahey

  • Author_Institution
    Department of Biomedical Engineering, Duke University, Durham, North Carolina, United States of America
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Conventional multiple track location shear wave elasticity imaging (MTL-SWEI) is a powerful tool for noninvasively estimating tissue elasticity. The resolution and noise of MTL-SWEI systems, however, are limited by ultrasound speckle. Single track location SWEI (STL-SWEI) is a novel variant which fixes the position of the tracking beam and modulates the push location to effectively cancel out the effects of speckle bias. We present a 3D STL-SWEI system which provides full suppression of lateral and elevation speckle bias for high resolution 3D elasticity imaging, and requires no inherent spatial filtering to make precise measurements of shear wave speed. The system is demonstrated in a uniform elasticity phantom.
  • Keywords
    "Imaging","Elasticity","Three-dimensional displays","Gratings","Speckle","Image resolution","Pistons"
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2015 IEEE International
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2015.0036
  • Filename
    7329264