• DocumentCode
    3704486
  • Title

    Development of a 27.12 MHz CC-LCF intraluminal applicator for hyperthermia of the esophagus

  • Author

    J Crezee;HJ Harms;HJ Boluijt;J Sijbrands;MCCM Hulshof;HP Kok

  • Author_Institution
    Dept. Radiation Oncology, Academic Medical Center, Amsterdam, The Netherlands
  • fYear
    2015
  • Firstpage
    769
  • Lastpage
    772
  • Abstract
    Esophageal tumors can be treated with radiation combined with hyperthermia. This study investigated the design of a 27.12 MHz capacitively coupled intraluminal applicator for hyperthermia of the esophagus to be used in combination with external locoregional heating. It consists of four independent electrodes mounted on an inflatable balloon catheter, 10mm each, Measured SAR penetration depth is 4.5mm. Spatial steering is excellent and will prove useful in heterogeneous tumors. Combination with locoregional heating showed a strong increase in temperature rise. Next step is application in esophageal cancer.
  • Keywords
    "Applicators","Hyperthermia","Electrodes","Temperature measurement","Tumors","Heating","Temperature distribution"
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference (EuMC), 2015 European
  • Type

    conf

  • DOI
    10.1109/EuMC.2015.7345877
  • Filename
    7345877