DocumentCode
3717962
Title
Development of prototype system for robot-assisted ultrasound diagnosis
Author
Joonho Seo;Jangho Cho;Hyunsoo Woo;Yongkoo Lee
Author_Institution
Department of Medical Assistive Robot, Korea Institute of Machinery &
fYear
2015
Firstpage
1285
Lastpage
1288
Abstract
This paper presents a prototype system for robot-assisted ultrasound diagnosis. The system includes master-slave configuration integrated with ultrasound diagnostic machine. We designed a new slave robot based on rotary Stewart platform and the conventional US probe is installed inside of the platform and contacts with patient´s skin through the hole in the base frame. In the experiments, 3D workspace including circular motion, tilting and twisting motion was investigated. Position accuracy between master device and slave robot was evaluated. Tele-robotic real-time US imaging was also tested by phantom model.
Keywords
"Robots","Medical diagnostic imaging","Three-dimensional displays","Prototypes","Yttrium"
Publisher
ieee
Conference_Titel
Control, Automation and Systems (ICCAS), 2015 15th International Conference on
ISSN
2093-7121
Type
conf
DOI
10.1109/ICCAS.2015.7364834