• DocumentCode
    3772828
  • Title

    Dose measurements at epithermal beams of research reactors with fricke gel and thermoluminescence detectors

  • Author

    G. Gambarini;E. Artuso;M. Felisi;D. Giove;S. Agosteo;L. Barcaglioni;L. Garlati;A. Pola;M. Borroni;M. Carrara;V. Klupak;L. Viererbl;M. Vins;M. Marek

  • fYear
    2015
  • fDate
    4/1/2015 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    5
  • Abstract
    Suitable dosimeter methods have to be developed to measure the different dose contributions in phantoms exposed to epithermal/thermal neutron beams of a research reactor. The method based on Fricke xylenol orange gel dosimeter in form of layers has shown to be very effective for achieving images of the various dose components in air or in phantoms exposed to epithermal/thermal neutron beams with very high fluence rate. Another useful method is based on the use of TLD-700 chips, from whose response the gamma dose and the thermal neutron fluence can be obtained by means of appropriate parameters of the glow curve.
  • Keywords
    "Neutrons","Inductors","Particle beams","Phantoms","Dosimetry","Standards","Photonics"
  • Publisher
    ieee
  • Conference_Titel
    Advancements in Nuclear Instrumentation Measurement Methods and their Applications (ANIMMA), 2015 4th International Conference on
  • Type

    conf

  • DOI
    10.1109/ANIMMA.2015.7465600
  • Filename
    7465600