DocumentCode
3775549
Title
Redesign of hydrogen isotope permeability measurement device for silicon-carbide and evaluation of minimum detectable flows at high temperature
Author
Yaushi Yamamoto;Hirosato Yamaguchi;Takehiro Yamamoto;Yuichiro Murakami;Daigo Yonetsu;Satoshi Konishi
Author_Institution
Faculty of Engineering Science, Kansai University, Suita, Osaka 564-8680, Japan
fYear
2015
fDate
5/1/2015 12:00:00 AM
Firstpage
1
Lastpage
5
Abstract
We have modified the vacuum exhaustion system in the permeation measurement device for Silicon-Carbide to reduce wraparound flows through sample holder. In the experiment, the pressure in the glass tube region decreases more than 20 times, and measured permeation flows also decreases about 1-order. The wraparound flow is estimated to decrease less than 1% of the permeation flows through SiC sample. It is evaluated that the deuterium permeation coefficient of SiC is about 100 times less than that of 316 stainless steel at 800 C.