• DocumentCode
    3805947
  • Title

    Neuroelectric Signal Analysis Using Nuclear Instrumentation Techniques

  • Author

    A. D. Carlson;B. Soucek

  • Author_Institution
    State University of New York at Stony Brook Stony Brook, New York
  • Volume
    18
  • Issue
    1
  • fYear
    1971
  • Firstpage
    41
  • Lastpage
    45
  • Abstract
    A real-time system for nerve spike recognition and separation has been developed. The system is based on the techniques, commonly used in nuclear instrumentation, which have been adapted here for biological signal measuring and processing. Techniques used are signal shaping and optimum noise filtering; pulse peak stretching; peak sampling; pulse shape discrimination and multichannel pulse height analysis. Spectra for nerve pulse heights of the cockroach are shown, for the case of spontaneous activity as well as for evoked activity (e. g. for a living object exposed to radiation wind or light). Similarities and contrasts between nuclear and biomedical signal analysis are discussed.
  • Keywords
    "Signal analysis","Instruments","Pulse shaping methods","Real time systems","Signal processing","Biomedical measurements","Nuclear measurements","Multi-stage noise shaping","Filtering","Signal sampling"
  • Journal_Title
    IEEE Transactions on Nuclear Science
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1971.4325837
  • Filename