    Validation of a multi-segment foot and ankle kinematic model for pediatric gait

  • Author

    Myers, K.A. ; Wang, M. ; Marks, R.M. ; Harris, G.F.

  • Author_Institution
    Dept. of Biomed. Eng., Marquette Univ., Milwaukee, WI, USA
  • Volume
    2
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    1497
  • Abstract
    This paper quantifies the system characteristics for a pediatric foot and ankle biomechanical model. While orientated along the Z-axis the static system resolution is computed at 0.32 ± 0.29 mm with 99.9% accuracy. Dynamic resolution and accuracy are 0.43 ± 0.39 mm and 99.8%, respectively. Angular dynamic resolution computes to 0.52 ± 3.36 degrees at 99.6% accuracy. These calculations are comparable to the Milwaukee adult foot and ankle model.
  • Keywords
    computer graphics; gait analysis; image motion analysis; kinematics; medical image processing; paediatrics; physiological models; Euler rotations; VICON motion analysis system; accurate model; angular dynamic resolution; biomechanics; dummy segment; foot and ankle kinematic model; linear testing; multisegment model; pediatric gait; stance; static system resolution; swing; Biomechanics; Biomedical computing; Biomedical engineering; Educational institutions; Foot; Ground support; Kinematics; Motion analysis; Orthopedic surgery; Pathology;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-7211-5
  • Type

    conf

  • DOI
    10.1109/IEMBS.2001.1020490
  • Filename