• DocumentCode
    3810702
  • Title

    Application of Hamamatsu S8550 APD Array to the Common PET/CT Detector

  • Author

    A. Nassalski;M. Moszynski;A. Syntfeld-Kzuch;L Swiderski;T. Szczesniak;D. Wolski;T. Batsch

  • Author_Institution
    Soltan Inst. for Nucl. Studies, Otwock
  • Volume
    55
  • Issue
    5
  • fYear
    2008
  • Firstpage
    2460
  • Lastpage
    2464
  • Abstract
    The progress of nuclear medicine and X-ray computer imaging encourage the researchers to develop new dual modality positron emission/X-rays tomography (PET/CT) systems, to acquire both the anatomical and the functional information during the same measurement, relying on the same detection system and electronics at the same time. The aim of this work was to study a detector consisting of four LSO pixel crystals coupled to a Hamamatsu S8550 avalanche photodiode (APD) array. The measurements covered tests of the detectors in the PET and in the CT modes using counting and current modes, respectively. The measurements included a determination of the light output, energy resolution, non-proportionality of the light yield and time resolution for 511 keV annihilation quanta, characterizing the PET detector, and the dependence of a counting rate versus a mean current of APD in the X-ray detection.
  • Keywords
    "Sensor arrays","Positron emission tomography","Computed tomography","X-ray detection","Time measurement","X-ray detectors","Current measurement","Energy resolution","Application software","Nuclear medicine"
  • Journal_Title
    IEEE Transactions on Nuclear Science
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2008.2002439
  • Filename
    4696575