Calibration of RHIC electron detectors
Author
He, P. ; Hseuh, H.C. ; Gassner, D. ; Gullotta, J. ; Trbojevic, D. ; Zhang, S.Y.
Author_Institution
Collider-Accelerator Dept., Brookhaven Nat. Lab., Upton, NY, USA
Volume
2
fYear
2003
fDate
12-16 May 2003
Firstpage
1425
Abstract
To characterize the electron cloud in RHIC, eleven custom electron detectors have been designed, fabricated and installed at a few RHIC warm-bore vacuum sectors for data collection during FY2003 runs. Prior to installation, the transmission and collection efficiencies of these detectors at various grid and collector bias voltages were measured using an electron gun with energy up to 1500 eV. This paper describes the design of the detector and the test system set-up, and summarizes the calibration results. In addition the calibration of a commercial microchannel plate is also reported.
Keywords
beam handling techniques; electron accelerators; electron detection; electron guns; ion accelerators; particle beam diagnostics; particle beam dynamics; storage rings; synchrotrons; RHIC electron detectors; RHIC warm-bore vacuum sectors; bias voltages; commercial microchannel plate; electron cloud; electron gun; Boring; Calibration; Clouds; Detectors; Electrons; Energy measurement; Gold; Particle beams; Testing; Voltage;