MR-tomography on patients with heart pacemakers-a numerical study
Author
Golombeck, M.-A. ; Dössel, O.
Author_Institution
Inst. of Biomed. Eng., Karlsruhe Univ., Germany
Volume
1
fYear
2004
fDate
1-5 Sept. 2004
Firstpage
1076
Lastpage
1079
Abstract
Patients having a heart pacemaker are not allowed to go to MR tomography (MRT). One of the most dangerous effects is the heating of the tissue around the electrode caused by the coupling to the RF field of the MR system. Experiments have been carried out using tanks filled with saline water and large heating has sometimes been observed. Other experiments e.g. with electrodes in the brain did not show any heating at all. In this work numerical studies have been carried out to understand the different results. In conclusion it is suggested that MRT could be possible if the "normal" geometry of the wires of a heart pacemaker is ensured and an open MR system is used.
Keywords
biological effects of fields; biological tissues; biomedical MRI; biomedical electrodes; brain; cardiology; numerical analysis; pacemakers; MR-tomography; RF field coupling; electrode; heart pacemaker; numerical study; saline water; tissue heating; Conductivity; Electrodes; Heart; Heating; Imaging phantoms; Magnetic resonance imaging; Pacemakers; Radio frequency; Temperature; Wire; MRT; Pacemaker;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-8439-3
Type
conf
DOI
10.1109/IEMBS.2004.1403350
Filename
1403350
Link To Document