Abstract
The main purpose of this development is to design an "intelligent sensor", composed of one multi anodes PMT, four custom front-end analog chips with embedded analog to digital converter and a pre-processing logical unit. The idea of this design is to improve as most as possible the treatment speed of signals of the PMT. This design must be compact to fit the PMT dimensions, in order to combine them together to form an assembly which will be used in medical imaging research, especially for gamma-ray camera at 144 keV (99mTc) with NaI(Tl) cristal.
Keywords
analogue-digital conversion; cameras; gamma-ray apparatus; intelligent sensors; medical image processing; mixed analogue-digital integrated circuits; nuclear electronics; solid scintillation detectors; 144 keV; ASIC; NaI(Tl) crystal; analog to digital converter; application specific integrated circuit; gamma-ray camera; intelligent sensors; medical imaging research; multianode PMT; Analog-digital conversion; Anodes; Application specific integrated circuits; Bandwidth; Cameras; Clocks; Field programmable gate arrays; Performance gain; Signal design; Universal Serial Bus;