    Comparison of analytic and iterative reconstruction methods for quantitative cardiac PET studies in 3D using Oxygen-15 water scans

  • Author

    Bouchareb, Yassine ; Thielemans, Kris ; Spinks, Terence ; Rimoldi, Ornella ; Camici, Paolo G.

  • Author_Institution
    Medical Res. Council, Clinical Sci. Center, London, UK
  • Volume
    4
  • fYear
    2005
  • fDate
    23-29 Oct. 2005
  • Firstpage
    2120
  • Lastpage
    2123
  • Abstract
    We investigated the quantitative accuracy of iterative and analytic reconstruction methods to determine myocardial blood flow (MBF) in PET studies in pigs using O-15 labeled water. Dynamic scans were acquired on the EXACT3D PET scanner on large white pigs after H215O injection (resting and dipyridamole-induced stress). Radioactive microspheres were injected during the H215O scans to provide a "gold standard" of MBF values. Data were reconstructed with 3DRP, FORE + attenuation-weighted OSEM, FORE-FBP and 3D-OSEM. Factor images were generated by cluster analysis and linear dimension reduction (myocardium and left and right cavities) and resliced to short axis images; 16 ROIs were defined in the left myocardium and 2 ROIs in the left and right cavities. ROIs were projected onto the dynamic images to extract time-activity-curves (TACs), which were then fitted to a single compartment model to estimate regional and global absolute MBF. Microsphere measurements were obtained in a similar way and 64 pairs of measurements were made. FORE-OSEM and 3D-RP overestimated the CFR values by 10% and 17%, respectively and 3D-OSEM and FORE-FBP underestimate CFR values by 3% and 17%, respectively. Regression analysis for 3D-OSEM and FORE-OSEM gave intercepts close to zero and slopes close to unity. Bland and Altman plots of differences between all reconstructions and Microsphere provided a mean difference and limits of agreements in favour of 3D-OSEM and FORE-OSEM (mean difference closer to zero and smaller SD). For absolute quantification of MBF using O-15 water, 3D-OSEM and FORE-OSEM proved to be at least as accurate as 3DRP and FORE-FBP compared with the Microsphere technique.
  • Keywords
    cardiology; expectation-maximisation algorithm; image reconstruction; medical image processing; positron emission tomography; 3D-OSEM; 3DRP; Altman plots; Bland plots; EXACT3D PET scanner; FORE-FBP; H215O injection; Microsphere technique; O-15 labeled water; Oxygen-15 water scans; analytic reconstruction; attenuation-weighted OSEM; cavities; cluster analysis; dipyridamole-induced stress; factor images; iterative reconstruction; large white pigs; linear dimension reduction; myocardial blood flow; myocardium; quantitative cardiac PET studies; radioactive microspheres; regression analysis; time-activity-curves; Blood flow; Gold; Image analysis; Image generation; Image reconstruction; Iterative methods; Myocardium; Positron emission tomography; Reconstruction algorithms; Stress;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2005 IEEE
  • ISSN
    1095-7863
  • Print_ISBN
    0-7803-9221-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2005.1596753
  • Filename
    1596753