• DocumentCode
    471574
  • Title

    Modeling and simulation of physical performance of a External Unilateral Mechatronic Orthopaedic Fixator - Bone system

  • Author

    Lesnicwska, A. ; Choromanski, W. ; Deszezynski, J. ; Dobrzynski, G.

  • Author_Institution
    Fac. of Transp., Warsaw Univ. of Technol.
  • fYear
    2006
  • fDate
    Aug. 30 2006-Sept. 3 2006
  • Firstpage
    1533
  • Lastpage
    1536
  • Abstract
    Restricted element study of the fracture healing by external fixation device was investigated. The analyses were performed under an axial and variable loaded boundary conditions. The effect of different fracture size and different distance between bone and the external fixator device on the stress distribution was investigated. The results show that stresses in the external fixator device are highest at the beginning of the fracture healing process, and are gradually decreasing with the time of the treatment. The analyses were carried out using the commercial package CATIA P3 V5R11. This allowed to build a three-dimensional model more similar to the geometrical architecture of the long bone as well as of the external fixator. Three-dimensional restricted element model also allowed a collection of more realistic results. However, the accuracy of the results depends not only on the quality of the model geometry but also on the material properties assigned to the model components. It also depends on the accuracy in the simulation of the finite element model and the optimized mesh generation
  • Keywords
    biomechanics; bone; fracture; kinematics; mesh generation; orthopaedics; orthotics; patient rehabilitation; CATIA P3 V5R11commercial package; axial loaded condition; bone; external fixator device; finite element model; fracture healing process; fracture size; optimized mesh generation; stress distributions; three-dimensional restricted element model; variable loaded boundary condition; Bones; Boundary conditions; Geometry; Material properties; Mechatronics; Orthopedic surgery; Packaging; Performance analysis; Solid modeling; Stress;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
  • Conference_Location
    New York, NY
  • ISSN
    1557-170X
  • Print_ISBN
    1-4244-0032-5
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2006.259328
  • Filename
    4462056