    The voltage-current characteristics of an electrosurgical arc

  • Author

    LaCourse, J.R. ; Rothwell, A.D. ; Selikowitz, S.M.

  • Author_Institution
    Biomed. Eng. Center, Univ. of New Hampshire Durham, Durham, NH, USA
  • Volume
    3
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    1136
  • Lastpage
    1137
  • Abstract
    The voltage-current characteristics resulting from the application of a high voltage, 500 kHz sinusoidal frequency via a electrosurgical electrode in a simulated TUR environment are presented. Arc responses are clearly seen in the tissue. Rectification is observed in the power density spectra. These results indicate that neuromuscular stimulation is possible.
  • Keywords
    biological tissues; biomedical electrodes; neuromuscular stimulation; rectification; surgery; electrosurgical arc; electrosurgical electrode; frequency 500 kHz; neuromuscular stimulation; power density spectra; rectification; simulated TUR environment; sinusoidal frequency; tissue; voltage-current characteristics; Biomedical measurements; Circuit faults; Current measurement; Electrodes; Frequency measurement; Frequency modulation; Voltage measurement;