Abstract
Extracorporeal circuits are used in artificial organs for blood treatment such as hemodialysis (HD), hemofiltration, plasmafiltration and ECMO (Extracorporeal Membrane Oxygenation). A malfunction of the circuit may cause blood loss to the environment. Safety standards therefore require protective systems which detect and prevent such blood losses. An accepted method is the use of pressure monitors. The paper de- scribes a fail-safe pressure monitor based on a simple algorithm utilizing the pulsations of roller pumps.