DocumentCode
55966
Title
Gas–Liquid Two-Phase Flow Correction Method for Digital CMF
Author
Qi-Li Hou ; Ke-Jun Xu ; Min Fang ; Yan Shi ; Bo-Bo Tao ; Rong-Wei Jiang
Author_Institution
Sch. of Electr. & Autom. Eng., Hefei Univ. of Technol., Hefei, China
Volume
63
Issue
10
fYear
2014
fDate
Oct. 2014
Firstpage
2396
Lastpage
2404
Abstract
There are large measurement errors in Coriolis mass flowmeter (CMF) when it is used to measure gas-liquid two-phase flow without compensation. In particular, the flowtube of CMF may stall when using analog transmitters and CMF cannot operate normally because of the high damping caused by gas-liquid two-phase. To solve above problems, a gas-liquid two-phase flow rig is developed to perform the two-phase flow experiments. A digital drive method using multiplying digital-to-analog converter and direct digital synthesizer is applied to CMF to maintain the flowtube oscillation under two-phase flow condition. A digital zero-crossing detection method based on Lagrange interpolation is adopted to calculate the frequency and phase difference of the sensor output signals in order to synthesize the digital drive signal and realize the flow measurement. The artificial neural network using back propagation algorithm is applied to correct measurement errors. A digital Coriolis mass flow transmitter is developed with a digital signal processor to control the digital drive, and realize the two-phase flow measurement and correction. By connecting the developed transmitter with a CMF025 type primary instrument, the gas-liquid two-phase flow experiments are conducted to validate the performance of the transmitter. The testing results show that the transmitter can maintain the flowtube oscillation and reduce the metering errors.
Keywords
backpropagation; computerised instrumentation; damping; digital signal processing chips; digital-analogue conversion; direct digital synthesis; flow sensors; flowmeters; frequency measurement; interpolation; mass measurement; measurement errors; neural nets; phase measurement; pipe flow; transmitters; two-phase flow; Coriolis mass flowmeter; Lagrange interpolation; analog transmitter; artificial neural network; back propagation algorithm; compensation; damping; digital CMF025 type primary instrument; digital Coriolis mass flow transmitter; digital drive method; digital signal processor; digital zero-crossing detection method; direct digital synthesizer; flowtube oscillation; frequency difference calculation; gas-liquid two-phase flow correction method; gas-liquid two-phase flow measurement; measurement error; multiplying digital-to-analog converter; phase difference calculation; sensor; Digital signal processing; Instruments; Interpolation; Liquids; Measurement errors; Oscillators; Transmitters; Artificial neural network; Coriolis mass flowmeter (CMF); digital drive method; digital zero-crossing detection method; gas??liquid two-phase flow;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2014.2310031