DocumentCode
640829
Title
Modernization and calibration of intrferometer of TOKAMAK “Compass”
Author
Varavin, M. ; Zajac, Jozef ; Zacek, F. ; Nanobashvili, S. ; Ermak, G. ; Varavin, A. ; Vasilev, Andrey ; Stumbra, M. ; Vetoshko, A. ; Fateev, A. ; Shevchenko, Valentin
Author_Institution
Assoc. EURAATOM/IPP.CR, Inst. of Plasma Phys., Prague, Czech Republic
fYear
2013
fDate
23-28 June 2013
Firstpage
571
Lastpage
575
Abstract
This article presents “Compass” TOKAMAK microwave diagnostics calibration and results accuracy enhancing experiment methodology. Main subject is microwave diagnostic system phase measurements description. Additional interest is mistakes, attendant to phase measurement. Emphasis is put on phase detectors measurements mistakes correction methodology; two different phase meter configurations are described. Phase detectors, based on AD8302 CPLD, results processing methodology are presented.
Keywords
Tokamak devices; calibration; phase detectors; phase measurement; phase meters; plasma diagnostics; radiowave interferometers; AD8302 CPLD; TOKAMAK compass; TOKAMAK microwave diagnostics calibration; accuracy enhancing experiment methodology; interferometer; microwave diagnostic system phase measurements description; modernization; phase detectors measurements mistakes correction methodology; phase meter configurations; Detectors; Frequency measurement; Phase measurement; Plasma density; Plasma measurements; Tokamaks;
fLanguage
English
Publisher
ieee
Conference_Titel
Physics and Engineering of Microwaves, Millimeter and Submillimeter Waves (MSMW), 2013 International Kharkov Symposium on
Conference_Location
Kharkiv
Print_ISBN
978-1-4799-1066-3
Type
conf
DOI
10.1109/MSMW.2013.6622126
Filename
6622126
Link To Document