DocumentCode
651455
Title
Multiplexed detection of waterborne pathogens with an array of microfluidic integrated high-sensitivity organic photodiodes
Author
Pires, N.M.M. ; Tao Dong
Author_Institution
Dept. of Micro & Nano Syst. Technol., Vestfold Univ. Coll., Tonsberg, Norway
fYear
2013
fDate
Oct. 31 2013-Nov. 2 2013
Firstpage
105
Lastpage
108
Abstract
It is reported a high-throughput microfluidic system integrating organic photodetectors based on poly(2,7-carbazole) semiconducting polymer. The system reported here showed detection limits of 105-106 cells/mL for the chemiluminescence detection of two bacterial cells. The detection of viral particles as low as 10-4 μg/mL was also shown. High specificity and ability to perform multiplexed analysis in an integrated lab-on-a-chip device was demonstrated.
Keywords
bioMEMS; biological techniques; cellular biophysics; chemiluminescence; lab-on-a-chip; microfluidics; microorganisms; organic semiconductors; photodetectors; photodiodes; polymers; bacterial cells; chemiluminescence detection; high-throughput microfluidic system; integrated lab-on-a-chip device; microfluidic integrated high-sensitivity organic photodiode array; multiplexed analysis; multiplexed detection; organic photodetectors; poly(2,7-carbazole) semiconducting polymer; viral particle detection; waterborne pathogens; Arrays; Detectors; Microfluidics; Multiplexing; Pathogens; Photoconductivity; Photodiodes;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Circuits and Systems Conference (BioCAS), 2013 IEEE
Conference_Location
Rotterdam
Type
conf
DOI
10.1109/BioCAS.2013.6679650
Filename
6679650