Error analysis and calibration of two-way microwave time transfer equipment
Author
Fu Yongjie ; Yang Zhenji
Author_Institution
Key Defence Sci. & Technol. Lab. for Fuze Dynamic Characteristics, Beijing Inst. of Technol. China, Beijing, China
Volume
1
fYear
2013
fDate
16-19 Aug. 2013
Firstpage
483
Lastpage
487
Abstract
Through Analyzing the advantages and disadvantages of two-way satellite time transfer (TWSTT), two-way time microwave transfer and two-way optical fiber time transfer, selects two-way microwave time transfer method as time synchronization method of line-of-sight range. It introduces the basic principle of two-way microwave time transfer, using mathematical formula deduces error sources. The influence of ground equipment on the synchronous precision are analyzed in detail, and describes the way that reduces the modem delay and the calibration. After analyzing the synchronous precision of two-way microwave time, it can reach ns magnitude, and this method is satisfied to the requirements of multi-radar and ships in the time synchronization precision.
Keywords
calibration; delays; error analysis; microwave measurement; time measurement; calibration; error analysis; ground equipment; line-of-sight range; time synchronization method; two-way microwave time transfer equipment; two-way optical fiber time transfer; two-way satellite time transfer; Delays; Microwave FET integrated circuits; Microwave communication; Microwave integrated circuits; Modems; Temperature measurement; Error analysis; calibration; time and frequency; two-way microwave time transfer;
fLanguage
English
Publisher
ieee
Conference_Titel
Electronic Measurement & Instruments (ICEMI), 2013 IEEE 11th International Conference on
Conference_Location
Harbin
Print_ISBN
978-1-4799-0757-1
Type
conf
DOI
10.1109/ICEMI.2013.6743088
Filename
6743088
Link To Document