DocumentCode
708831
Title
In situ and ex vivo MPI performance compared to fluorescent and MRI imaging
Author
Bo Zheng ; von See, Marc ; Kuan Lu ; Goodwill, Patrick ; Conolly, Steven
fYear
2015
fDate
26-28 March 2015
Firstpage
1
Lastpage
1
Abstract
This paper shows the first direct comparisons between magnetic particle imaging (MPI) and other standard preclinical imaging modalities (computerised tomography, optical imaging, magnetic resonance imaging) for image contrast and depth attenuation. Compared to optical imaging and MRI, MPI signals are not affected by tissue surrounding the probe, making MPI potentially a more quantitative and robust molecular imaging technique.
Keywords
biological tissues; biomedical MRI; biomedical optical imaging; computerised tomography; fluorescence; molecular biophysics; MRI imaging; biological tissues; computerised tomography; depth attenuation; ex vivo MPI; fluorescent imaging; image contrast; in situ MPI; magnetic resonance imaging; molecular imaging; optical imaging; Attenuation; Computed tomography; Fluorescence; Magnetic resonance imaging; Optical imaging; Probes;
fLanguage
English
Publisher
ieee
Conference_Titel
Magnetic Particle Imaging (IWMPI), 2015 5th International Workshop on
Conference_Location
Istanbul
Print_ISBN
978-1-4799-7269-2
Type
conf
DOI
10.1109/IWMPI.2015.7106999
Filename
7106999
Link To Document