• DocumentCode
    731977
  • Title

    Encapsulation of integrated circuits in plastic microfluidic systems using hot embossing

  • Author

    Iyer, V. ; Murali, P. ; Paredes, J. ; Liepmann, D. ; Boser, B.

  • Author_Institution
    Berkeley Sensors & Actuators Center, Berkeley, CA, USA
  • fYear
    2015
  • fDate
    21-25 June 2015
  • Firstpage
    1822
  • Lastpage
    1825
  • Abstract
    A method for fabricating thermoplastic cartridges encapsulating integrated circuits, electrodes, and microfluidic channels using hot embossing is described for the application of a point of care magnetic label flow cytometer. Finished devices with a microchannel depth of 86 μm and RMS surface roughness of 1.6 μm are capable of operating at pressures of up to 227 kPa without leaks. The device was used to evaluate the efficiency of an on chip magnetic focusing structure which successfully focused 98% of the 4.5 μm Dynabeads® to a 30 μm target region using 100 mA of current and a 4 mT external field.
  • Keywords
    embossing; encapsulation; integrated circuit packaging; microfluidics; plastic products; Dynabeads; care magnetic label flow cytometer; current 100 mA; electrodes; hot embossing; integrated circuits; magnetic flux density 4 mT; microfluidic channels; on chip magnetic focusing structure; plastic microfluidic systems; pressure 227 kPa; size 1.6 mum; size 30 mum; size 4.5 mum; size 86 mum; thermoplastic cartridges; Embossing; Fabrication; Focusing; Microchannels; Microfluidics; Plastics; Sensors; CMOS; Microfluidics; flow cytometer; hot-embossing; lab-on-achip; magnetophoresis; packaging; plastic;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS), 2015 Transducers - 2015 18th International Conference on
  • Conference_Location
    Anchorage, AK
  • Type

    conf

  • DOI
    10.1109/TRANSDUCERS.2015.7181302
  • Filename
    7181302