• DocumentCode
    747989
  • Title

    Accuracy of an atomic microwave power standard

  • Author

    Paulusse, David C. ; Rowell, Nelson L. ; Michaud, Alain

  • Author_Institution
    Inst. for Nat. Meas. Stand., Nat. Res. Council, Ottawa, Ont., Canada
  • Volume
    54
  • Issue
    2
  • fYear
    2005
  • fDate
    4/1/2005 12:00:00 AM
  • Firstpage
    692
  • Lastpage
    695
  • Abstract
    We have built an atomic microwave power standard based on the electromagnetic interaction with laser-cooled atoms. The atoms traversed a waveguide transmission line, and under the effect of the radiation, the internal state populations underwent a Rabi flopping oscillation. Measurement of the oscillation frequency allowed the determination of the incident microwave power. As many of 60 oscillations were observed over a dynamic range of 20 dB and the standard deviation of the measurements was about 0.02%. The measured frequency was compared to a calculated one and an agreement of 1.3% with an uncertainty of 5% (rectangular) was found.
  • Keywords
    laser cooling; measurement standards; measurement uncertainty; microwave measurement; power measurement; radiation effects; Rabi flopping oscillation; atomic measurements; atomic microwave power standard; electromagnetic interaction; incident microwave power; laser cooling; laser-cooled atoms; measurement uncertainty; microwave measurements; oscillation frequency measurement; radiation effect; waveguide transmission line; Atom lasers; Atomic beams; Atomic measurements; Electromagnetic radiation; Electromagnetic waveguides; Frequency measurement; Masers; Power lasers; Power transmission lines; Waveguide lasers; Atomic measurements; cooling; laser cooling; microwave measurements;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2005.843537
  • Filename
    1408266