• DocumentCode
    757213
  • Title

    Small Size Integrated CsI(Tl) Spectrometer Efficiency and Properties Dependence on Temperature

  • Author

    Dolev, Eli ; Manor, Avi ; Brandys, Irad ; Tirosh, Danny ; Ziskind, Gennady ; Orion, Itzhak

  • Author_Institution
    Ben-Gurion Univ. of the Negev, Beer-Sheva
  • Volume
    55
  • Issue
    3
  • fYear
    2008
  • fDate
    6/1/2008 12:00:00 AM
  • Firstpage
    1237
  • Lastpage
    1240
  • Abstract
    Small-sized radiation scintillation detectors have been extensively developed for portable radiation survey systems, for a variety of applications. The CsI(Tl) capacity to discriminate the energy-lines of a gamma-radiation source makes these detectors suitable for personal detection and spectrometry. Additional advantages are their miniature electronics and operation at room temperature. However, it is essential to inspect their efficiency, sensitivity and stability under the practical range of the environmental conditions expected during operation. The aim of this research is to study the changes in the response of a CsI(Tl) detector, 22 mm in diameter and 25 mm long, in the full range of operation temperatures, from -20 to 60 degC . Results show that the photopeak efficiency for the Cs-137 gamma-line remains constant, even though the peak center channel is shifted as function of the detector temperature. The peak spectral resolution was also analyzed in the same temperature range and was found to be around 8%. Since the photopeak efficiency has been constant over the entire tested temperature range, the CsI(Tl) detector is a suitable spectrometry detector for environmental operations.
  • Keywords
    gamma-ray detection; gamma-ray spectrometers; nuclear electronics; photomultipliers; radioactive sources; solid scintillation detectors; 137Cs gamma-line; CsI(Tl) spectrometer; environmental conditions; gamma-radiation source; miniature electronics; personal radiation detector; photomultiplier tube; portable radiation survey systems; room temperature; size 22 mm; small-sized radiation scintillation detectors; spectral resolution; temperature -20 C to 60 C; Gamma ray detection; Gamma ray detectors; Scintillation counters; Solid scintillation detectors; Spectroscopy; Stability; Temperature dependence; Temperature distribution; Temperature sensors; Testing; CsI(Tl); gamma-ray spectrometry; inorganic scintillation detector; temperature dependence;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2008.922843
  • Filename
    4545107