• DocumentCode
    782557
  • Title

    A reflected-scanned ultrasound system for external simultaneous thermoradiotherapy

  • Author

    Moros, Eduardo G. ; Straube, William L. ; Myerson, Robert J.

  • Author_Institution
    Mallinckrodt Inst. of Radiol., Washington Univ. Sch. of Med., St. Louis, MO, USA
  • Volume
    43
  • Issue
    3
  • fYear
    1996
  • fDate
    5/1/1996 12:00:00 AM
  • Firstpage
    441
  • Lastpage
    449
  • Abstract
    The simultaneous delivery of hyperthermia and ionizing radiation has the potential to improve clinical outcome. To this purpose, a scanning ultrasound reflector-linear array system (SURLAS) with the ability both to conform power to superficial volumes and to operate concomitantly with medical linear accelerators is currently under development. In this purpose-specific design, the ultrasound waves generated by a linear array are directed toward a scanning reflector which in turn deflects the waves toward the target. In previous experiments, the technical feasibility of this design was demonstrated. Here, the authors are concerned with the minimization of a key design parameter, namely, the array element size, in order to minimize the amount of attenuating/scattering water-equivalent medium that a photon or a electron beam passes through before entering the target. First, the SURLAS design is described. Second, an acoustic model to compute power deposition patterns is presented. This model is coupled to a bioheat transfer model for computation of temperature fields. Third, an analysis is performed to determine the minimum array element size for three target categories. Finally, acoustic fields and temperature distributions induced by the SURLAS for the three target categories are presented. The analysis and simulations show that the SURLAS has the potential to induce uniform temperature distributions in large superficial volumes with small enough elements to allow simultaneous delivery with electron beam therapy.
  • Keywords
    biomedical equipment; biomedical ultrasonics; hyperthermia; radiation therapy; ultrasonic equipment; acoustic fields; clinical outcome improvement; design technical feasibility; electron beam therapy; external simultaneous thermoradiotherapy; key design parameter minimization; large superficial volumes; linear array-generated ultrasound waves; medical instrumentation; medical linear accelerators; minimum array element size; reflected-scanned ultrasound system; uniform temperature distributions; Acoustic scattering; Computational modeling; Electromagnetic scattering; Electron beams; Hyperthermia; Ionizing radiation; Linear accelerators; Particle scattering; Temperature distribution; Ultrasonic imaging;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.489403
  • Filename
    489403