DocumentCode
784724
Title
Calibration of an electronically scanned white-light interferometric transducer for measurements of high pressure
Author
Huang, Shi ; Morawski, Roman Z. ; Barwicz, Andrzej ; Bock, Wojtek J. ; Urbanczyk, W.
Author_Institution
Dept. d´´Ingenierie, Quebec Univ., Trois-Rivieres, Que., Canada
Volume
44
Issue
3
fYear
1995
fDate
6/1/1995 12:00:00 AM
Firstpage
698
Lastpage
703
Abstract
The paper deals with an electronically scanned white-light interferometric transducer for high-pressure measurements which consists of a sensor, a conditioning circuit, an analog-to-digital converter, and a digital signal processor of the output signal from the sensor. The calibration of such a transducer consists of identifying a function modeling the relationship between the raw result of measurement y, which is the output of the transducer, and the measured pressure p, taking into account an influence quantity-temperature T. An algorithm for calibration is proposed and studied
Keywords
CCD image sensors; calibration; light interferometers; light interferometry; pressure sensors; pressure transducers; quantisation (signal); signal processing equipment; transducers; CCD camera; analog-to-digital converter; calibration; conditioning circuit; digital signal processor; electronically scanned white-light interferometric transducer; function modeling; high-pressure measurements; sensor; Calibration; Charge coupled devices; Charge-coupled image sensors; Circuits; Interference; Optical fiber devices; Optical fiber polarization; Pressure measurement; Sensor phenomena and characterization; Transducers;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.387312
Filename
387312
Link To Document