    Charge Collection Efficiencies for Lithium-Drifted Silicon and Germanium Detectors in the X-Ray Energy Region

  • Author

    Zulliger, H.R. ; Aitken, D.W.

  • Author_Institution
    Department of Physics and High Energy Physics Laboratory Stanford University Stanford, California
  • Volume
    15
  • Issue
    1
  • fYear
    1968
  • Firstpage
    466
  • Lastpage
    474
  • Abstract
    Theoretical calculations showing the charge collection efficiencies of Ge(Li) and Si(Li) planar detectors as a function of trapping lengths and incident X-ray energies are presented. It is demonstrated that there should be an intrinsic energy dependence of the charge collection efficiency for radiation lengths comparable to the detector depletion depth. The nonlinear behavior is generally greater for unequal hole and electron trapping lengths. Experimental results confirm this in part, but surface effects appear to accentuate the effect. It is concluded that, neglecting surface effects, the nonlinearities should be eliminated if the radiation is incident in a direction perpendicular to the field. Highest charge collection efficiency, best resolution and linearity are obtained by injecting the photons through the detector side (perpendicular to the field) along a unique plane which is defined by the electron to hole trapping ratio.
  • Keywords
    Charge carrier processes; Electromagnetic wave absorption; Electron traps; Energy measurement; Germanium; Linearity; Radiation detectors; Silicon; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1968.4324888
  • Filename
    4324888