    Operational Characteristics of Germanium Detectors at Higher Temperatures

  • Author

    Pehl, Richard H. ; Haller, Eugene E. ; Cordi, Richard C.

  • Author_Institution
    Lawrence Berkeley Laboratory University of California Berkeley, California 94720
  • Volume
    20
  • Issue
    1
  • fYear
    1973
  • Firstpage
    494
  • Lastpage
    499
  • Abstract
    Several important areas for the use of germanium spectrometers would be opened if these detectors could be successfully operated at temperatures considerably in excess of 77°K. Precise measurements on a series of detectors made from different high-purity germanium crystals and on one lithium-drifted germanium detector indicate that the resolution does not degrade significantly until nearly 150°K is reached. Similar-sized detectors, including the lithium-drifted device, exhibited essentially identical temperature-resolution relationships.
  • Keywords
    Crystals; Current measurement; Detectors; Diodes; Extraterrestrial measurements; Germanium; Leakage current; Steel; Temperature; Thermal resistance;