• DocumentCode
    809090
  • Title

    The status of the ICR/RAL BaF2-TMAE positron camera

  • Author

    Wells, K. ; Ott, R.J. ; Suckling, J. ; Bateman, J.E. ; Stephenson, R. ; Connolly, J.F.

  • Author_Institution
    R. Marsden Hospital, Belmont, UK
  • Volume
    39
  • Issue
    5
  • fYear
    1992
  • fDate
    10/1/1992 12:00:00 AM
  • Firstpage
    1475
  • Lastpage
    1479
  • Abstract
    A new prototype BaF2-TMAE (tetrakis dimethyl-aminoethylene) detector has been developed for use in a positron camera. Results of projection imaging are presented. The projection images obtained with the 100-cm2 sensitive area detector represent the first occasion a distributed source has been used in conjunction with a BaF2-TMAE detector for positron imaging. This is also the first result of 5-ns timing, 5-mm detector resolution and an efficiency of 20%, with the same chamber operating conditions. The chamber can operate in a stable condition for periods of several hours
  • Keywords
    barium compounds; biomedical equipment; radioisotope scanning and imaging; 5 ns; BaF2; TMAE; diagnostic nuclear medicine; distributed source; positron camera; positron imaging; projection images; tetrakis dimethyl-aminoethylene; Cameras; Circuits; Crystals; Detectors; Electrons; Insulation; Positrons; Spatial resolution; Vacuum systems; Wire;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.173227
  • Filename