• DocumentCode
    81
  • Title

    On the precision of time-of-flight shear wave speed estimation in homogeneous soft solids: initial results using a matrix array transducer

  • Author

    Wang, Michael ; Byram, Brett ; Palmeri, Mark ; Rouze, Ned ; Nightingale, Kathryn

  • Author_Institution
    Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
  • Volume
    60
  • Issue
    4
  • fYear
    2013
  • fDate
    Apr-13
  • Firstpage
    758
  • Lastpage
    770
  • Abstract
    A system capable of tracking radiation-force-induced shear wave propagation in a 3-D volume using ultrasound is presented. In contrast to existing systems, which use 1-D array transducers, a 2-D matrix array is used for tracking shear wave displacements. A separate single-element transducer is used for radiation force excitation. This system allows shear wave propagation in all directions away from the push to be observed. It is shown that for a limit of 64 tracking beams, by placing the beams at the edges of the measurement region of interest (ROI) at multiple directions from the push, time-of- flight (TOF) shear wave speed (SWS) measurement uncertainty can theoretically be reduced by 40% compared with equally spacing the tracking beams within the ROI along a single plane, as is typical when using a 1-D array for tracking. This was verified by simulation, and a reduction of 30% was experimentally observed on a homogeneous phantom. Analytical expressions are presented for the relationship between TOF SWS measurement uncertainty and various shear wave imaging parameters. It is shown that TOF SWS uncertainty is inversely proportional to ROI size, and inversely proportional to the square root of the number of tracking locations for a given distribution of beam locations relative to the push. TOF SWS uncertainty is shown to increase with the square of the SWS, indicating that TOF SWS measurements are intrinsically less precise for stiffer materials.
  • Keywords
    biomedical ultrasonics; elastic waves; ultrasonic imaging; 1D array transducers; 2D matrix array; 3D volume; beam locations; homogeneous phantom; homogeneous soft solids; matrix array transducer; radiation force excitation; radiation-force-induced shear wave propagation; shear wave displacements; shear wave imaging parameters; single-element transducer; time-of- flight shear wave speed measurement uncertainty; time-of-flight shear wave speed estimation; tracking beams; tracking locations; Computer Simulation; Image Processing, Computer-Assisted; Phantoms, Imaging; Signal-To-Noise Ratio; Transducers; Ultrasonography;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2013.2624
  • Filename
    6489810