DocumentCode
816781
Title
A β-γ coincidence system for environmental 131
Author
Paperiello, C.J. ; Matuszek, J.M.
Author_Institution
Radiological Sciences Laboratory Division of Laboratories and Research New York State Department of Health Albany, New York 12201
Volume
22
Issue
1
fYear
1975
Firstpage
642
Lastpage
644
Abstract
A β-γ coincidence system using a thin plastic scintillator and a NaI(Tl) well crystal has been constructed for 131I environmental measurements. A background of 0.01 cpm with a system efficiency of 11% has been achieved. The detection limit is 0.08 pCi/l for a 100-minute counting period and otherwise usual conditions.
Keywords
Costs; Dairy products; Detectors; Laboratories; Optical amplifiers; Optical coupling; Photomultipliers; Plastics; Pulse amplifiers; Spectroscopy;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1975.4327720
Filename
4327720
Link To Document