• DocumentCode
    824103
  • Title

    Performance and Characterization of New Micromachined High-Frequency Linear Arrays

  • Author

    Lukacs, Marc ; Yin, Jianhua ; Pang, Guofeng ; Garcia, Richard C. ; Cherin, Emmanuel ; Williams, Ross ; Mehi, Jim ; Foster, F. Stuart

  • Author_Institution
    Imaging Res., Toronto Univ., Ont.
  • Volume
    53
  • Issue
    10
  • fYear
    2006
  • fDate
    10/1/2006 12:00:00 AM
  • Firstpage
    1719
  • Lastpage
    1729
  • Abstract
    A new approach for fabricating high frequency (>20 MHz) linear array transducers, based on laser micromachining, has been developed. A 30 MHz, 64-element, 74-mum pitch, linear array design is presented. The performance of the device is demonstrated by comparing electrical and acoustic measurements with analytical, equivalent circuit, and finite-element analysis (FEA) simulations. All FEA results for array performance have been generated using one global set of material parameters. Each fabricated array has been integrated onto a flex circuit for case of handling, and the flex has been integrated onto a custom printed circuit board test card for ease of testing. For a fully assembled array, with an acoustic lens, the center frequency was 28.7 MHz with a one-way -3 dB and -6 dB bandwidth of 59% arid 83%, respectively, arid a -20 dB pulse width of -99 ns. The per-element peak acoustic power, for a plusmn30 V single cycle pulse, measured at the 10 mm focal length of the lens was 590 kPa with a -6 dB directivity span of about 30 degrees. The worst-case total cross talk of the combined array and flex assembly is for nearest neighboring elements and was measured to have an average level -40 dB across the -6 dB bandwidth of the device. Any significant deviation from simulation can be explained through limitations in apparatus calibration and in device packaging
  • Keywords
    acoustic imaging; acoustic transducer arrays; design engineering; finite element analysis; micromachining; FEA simulations; acoustic lens; apparatus calibration; device bandwidth; device packaging; equivalent circuit; finite-element analysis; laser micromachining; linear array transducers; micromachined high-frequency linear arrays; peak acoustic power; printed circuit board test card; Acoustic measurements; Acoustic pulses; Acoustic testing; Assembly; Circuit analysis; Circuit testing; Field emitter arrays; Frequency; Lenses; Performance analysis;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2006.105
  • Filename
    4012857