• DocumentCode
    824929
  • Title

    2D imaging of ultrasound fields using CCD array to map output of Fabry-Perot polymer film sensor

  • Author

    Lamont, M. ; Beard, P.C.

  • Author_Institution
    Dept. of Med. Phys. & Bioeng., Univ. Coll. London, UK
  • Volume
    42
  • Issue
    3
  • fYear
    2006
  • Firstpage
    187
  • Lastpage
    189
  • Abstract
    A 2D ultrasound field imaging system that uses a CCD array to map the optical output of a planar Fabry-Perot polymer film sensing interferometer is described. This system enables ultrasound fields to be rapidly mapped over a 6.6×4.7 mm aperture with spatial and temporal resolutions of the order of 10 μm and 10 ns, respectively. The concept was successfully demonstrated by imaging the acoustic field at the focus of a pulsed 5 MHz PZT ultrasound transducer. The technique offers an alternative to piezoelectric-based methods for high-resolution biomedical and industrial ultrasonic imaging and field characterisation applications.
  • Keywords
    CCD image sensors; Fabry-Perot interferometers; biomedical imaging; piezoelectric transducers; polymer films; ultrasonic imaging; ultrasonic transducers; 10 micron; 10 ns; 2D ultrasound field imaging system; 5 MHz; CCD arrays; Fabry-Perot polymer film sensor; PZT ultrasound transducers; acoustic field imaging; biomedical ultrasonic imaging; high-resolution ultrasonic imaging; industrial ultrasonic imaging; planar Fabry-Perot polymer film sensing interferometer;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:20064135
  • Filename
    1593320