• DocumentCode
    825855
  • Title

    Intrinsic efficiency of germanium - a basis for calculating expected detector efficiency

  • Author

    Strauss, Michael G. ; Sherman, Irvin S. ; Cattelino, Mark J. ; Brenner, Raul ; Pehl, Richard H.

  • Author_Institution
    Argonne National Laboratory, Argonne, Illinois 60439
  • Volume
    24
  • Issue
    1
  • fYear
    1977
  • Firstpage
    88
  • Lastpage
    92
  • Abstract
    A method is presented whereby the intrinsic efficiency of Ge is utilized to calculate the expected peak efficiency of detectors having a wide range of sizes. The intrinsic efficiency of Ge, which is the probability for total absorption, was measured at 122 and 136 keV in Ge(Li) coaxial detectors and HPGe planar detectors having an effective thickness ranging from 5 to 50 mm. At 136 keV it is 64% for a thickness of 10 mm and 82% for 20 mm, after which it levels off reaching 89% at 50 mm. It is shown that the peak efficiency of a detector is a product of only the intrinsic efficiency and the solid angle, once losses due to edge escape and detector imperfections (surface channels and high dislocation densities) are determined. The absolute and relative [to NaI(T1)] peak efficiency of a sample detector, calculated on the basis of intrinsic efficiency, are in good agreement with measured values. This method should find applications in the design of new detector systems particularly those for diagnostic imaging with 99Tc (140 keV).
  • Keywords
    Absorption; Coaxial components; Detectors; Germanium; Laboratories; Lead; Material properties; Solids; Spectroscopy; Thickness measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1977.4328648
  • Filename
    4328648