  • Abstract
    Wilson (1991) presented an ultrasonic wideband estimator for axial blood flow velocity estimation through the use of the 2-D Fourier transform. It was shown how a single velocity component was concentrated along a line in the 2-D Fourier space, where the slope was given by the axial velocity. Later, it was shown that this approach could also be used for finding the lateral velocity component by also including a lateral sampling. A single velocity component would then be concentrated along a plane in the 3-D Fourier space, tilted according to the 2 velocity components. This paper presents 2 new velocity estimators for finding both the axial and lateral velocity components. The estimators essentially search for the plane in the 3- D Fourier space, where the integrated power spectrum is largest. The first uses the 3-D Fourier transform to find the power spectrum, while the second uses a minimum variance approach. Based on this plane, the axial and lateral velocity components are estimated. Several phantom measurements, for flow-to-depth angles of 60, 75, and 90 degrees, were performed. Multiple parallel lines were beamformed simultaneously, and 2 different receive apodization schemes were tried. The 2 estimators were then applied to the data. The axial velocity component was estimated with an average standard deviation below 2.8% of the peak velocity, while the average standard deviation of the lateral velocity estimates was between 2.0% and 16.4%. The 2 estimators were also tested on in vivo data from a transverse scan of the common carotid artery, showing the potential of the vector velocity estimation method under in vivo conditions.
  • Keywords
    Fourier transforms; biomedical ultrasonics; blood vessels; haemodynamics; phantoms; spectral analysis; 2-D Fourier transform; 2-D vector velocity estimation; 3-D Fourier space; axial blood flow velocity estimation; common carotid artery; multidimensional spectrum analysis; phantom; ultrasonic wideband estimator; Blood flow; Fourier transforms; Goniometers; Imaging phantoms; In vivo; Multidimensional systems; Performance evaluation; Sampling methods; Testing; Ultra wideband technology; Adult; Blood Flow Velocity; Carotid Arteries; Computer Simulation; Humans; Image Enhancement; Image Interpretation, Computer-Assisted; Male; Models, Cardiovascular; Pilot Projects; Rheology; Ultrasonography;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2008.859
  • Filename
    4589188