• DocumentCode
    845712
  • Title

    An international comparison of power meter calibrations conducted in 1987

  • Author

    Moore, William J M ; So, Eddy ; Oldham, Nile M. ; Miljanic, Etar N. ; Bergeest, Rainer

  • Author_Institution
    Nat. Res. Council of Canada, Ottawa, Ont., Canada
  • Volume
    38
  • Issue
    2
  • fYear
    1989
  • fDate
    4/1/1989 12:00:00 AM
  • Firstpage
    395
  • Lastpage
    401
  • Abstract
    The results of an intercomparison of power meter calibrations conducted during 1987 between the National Research Council of Canada, Ottawa, the US National Bureau of Standards, Gaithersburg, the Physikalisch-Technische Bundesanstalt, Braunschweig, and the Institut Mihailo Pupin, Belgrade, are described. The comparison was implemented by a transfer standard consisting of a time-division multiplier watt-converter developed at the Institut Mihailo Pupin. The measurements were made at 120 V, 5 A, and 50 and 60 Hz, at power factors of 1.0, 0.5 lead and lag, and 0.0 lead and lag. An agreement between the laboratories of better than 20 ppm is obtained
  • Keywords
    calibration; power measurement; 120 V; 1987; 5 A; 50 Hz; 60 Hz; Institut Mihailo Pupin; National Research Council of Canada; Physikalisch-Technische Bundesanstalt; US National Bureau of Standards; international comparison; power meter calibrations; time-division multiplier watt-converter; transfer standard; Calibration; Councils; Instruments; Laboratories; NIST; Power measurement; Reactive power; Standards development; Switches; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.192314
  • Filename
    192314