• DocumentCode
    848087
  • Title

    Lead Oxide Glass Tubing Converters for Gamma Detection in MWPC

  • Author

    Lum, G.K. ; Green, M.I. ; Perez-Mendez, V. ; Tam, K.C.

  • Author_Institution
    Lawrence Berkeley Laboratory, Berkeley and Radiology Department, U.C. San Francisco
  • Volume
    27
  • Issue
    1
  • fYear
    1980
  • Firstpage
    157
  • Lastpage
    165
  • Abstract
    PbO glass converters for the detection of gamma rays with MWPC have been investigated. The converters were made by fusing glass tubes stacked to form a close-packed array and were treated in a H2 reduction process to provide a suitable electric drift field. Electron conversion efficiency, drift velocity, and spatial resolution of these arrays were investigated. Using 511 keV gamma rays, an electron conversion efficiency of 8% was established for the converter of 80% PbO by weight (1.4 mm inner diameter, 0.1 mm wall thickness and 2 cm thick).
  • Keywords
    Conductivity; Electrons; Gamma ray detection; Gamma ray detectors; Gamma rays; Glass; Hydrogen; Ionization; Lead compounds; Solids;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1980.4330820
  • Filename