DocumentCode
848643
Title
Data Acquisition, Reconstruction and Display for the Donner 280-Crystal Positron Tomograph
Author
Huesman, R.H. ; Cahoon, J.L.
Author_Institution
Donner Laboratory University of California Berkeley, CA 94720
Volume
27
Issue
1
fYear
1980
Firstpage
474
Lastpage
478
Abstract
The imaging system for the Donner 280-Crystal Positron Tomograph is described. The architecture of the hardwired data acquisition system (histogrammer) allows a data rate in excess of two million events/second while simultaneously correcting for accidental events. The histogrammer has eight memory cards to allow separate accumulation of time-slices of the cardiac cycle, or multiple-buffering for rapid sequential studies. Transfer time from the double-bussed histogrammer to a disk of the host computer system is less than one second. A hardwired device is able to reconstruct a 256Ã256 image from 140 angles in less than three seconds, but since calibration and correction for attenuation are presently performed in software, this time is increased to about 10 seconds. Images are displayed on a 256Ã256 raster scan display system with image manipulation capabilities.
Keywords
Computed tomography; Computer architecture; Computer displays; Computer peripherals; Control systems; Data acquisition; Histograms; Image reconstruction; Positrons; Software performance;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1980.4330871
Filename
4330871