• DocumentCode
    849700
  • Title

    Intravenous Angiography Using Computerized Fluoroscopy

  • Author

    Shaw, C.G. ; Ergun, D.L. ; Kruger, R.A. ; Mistretta, C.A. ; Crummy, A.B. ; Myerowitz, D. ; Strother, C.M. ; Sackett, J. ; Lysel, M. Van ; Zarnstorff, W. ; Turnipseed, W.

  • Volume
    27
  • Issue
    3
  • fYear
    1980
  • fDate
    6/1/1980 12:00:00 AM
  • Firstpage
    1042
  • Lastpage
    1046
  • Abstract
    During the past several years, we have developed dedicated digital algorithms for real time processing of X-ray transmission imaging data obtained from an image intensifier television fluoroscopy system. Three imaging modes called Mask Mode Fluoroscopy, Mask Mode Radiography, and Time Interval Difference (T. I. D.) Imaging have been tested in over 100 patients since installation of this. apparatus at the University of Wisconsin Clinical Science Center. Feasibility studies have been conducted to determine the capability of these techniques for imaging the entire cardiovascular system, including peripheral arteries, using intravenous injections of iodinated contrast material.
  • Keywords
    Angiography; Arteries; Cardiovascular system; Diagnostic radiography; Image intensifiers; Optical imaging; Real time systems; TV; Testing; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1980.4330975
  • Filename
    4330975