    Performance of Room Temperature Mercuric Iodide (Hgi2) Detectors in the Ultralow-Energy X-Ray Region

  • Author

    Dabrowski, Andrzej J. ; Iwanczyk, Jan S. ; Barton, Jeffrey B. ; Huth, Gerald C. ; Whited, Richard ; Ortale, Carol ; Economou, Thanasis E. ; Turkevich, Anthony L.

  • Author_Institution
    University of Southern California Medical Imaging Science Group 4676 Admiralty Way, Suite 932, Marina Del Rey, CA 90291
  • Volume
    28
  • Issue
    1
  • fYear
    1981
  • Firstpage
    535
  • Lastpage
    540
  • Abstract
    Experiments have been done to study the performance of mercuric iodide (HgI2) detectors in the ultralow-energy x-ray region. Energy resolution values of 245 eV (FWHM) for the Mg K¿ x-ray line at 1.25 keV and 225 eV (FWHM) for the electronic noise linewidth have been obtained for an HgI2 detector with painted carbon contacts using a pulsed-light feedback preamplifier; the whole system was operated at room temperature. The resolution values in the ultralow-energy region are still limited by electronic noise of the system. In an attempt to minimize x-ray attenuation in the front contact, detectors were prepared with thin evaporated Pd contacts. These detectors show a pronounced low-energy tailing of the photopeak below a few keV, in contrast to the spectra obtained by detectors with carbon contact. An attempt has been made to explain the tailing effect starting with models wich have been proposed to describe similar effects in Ge detectors.
  • Keywords
    Energy resolution; Feedback; Gamma ray detection; Gamma ray detectors; Photonic band gap; Preamplifiers; Spectroscopy; Temperature; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1981.4331234
  • Filename
    4331234