• DocumentCode
    852396
  • Title

    Neutron Damage in Ge(Hp) Coaxial Detectors

  • Author

    Darken, Lawrence S., Jr. ; Trammell, Rex C. ; Raudorf, Thomas W. ; Pehl, Richard H.

  • Author_Institution
    Solid State Division, Oak Ridge National Laboratory, Oak Ridge, TN 37830 and EG&G ORTEC, 100 Midland Road, Oak Ridge, TN 37830
  • Volume
    28
  • Issue
    1
  • fYear
    1981
  • Firstpage
    572
  • Lastpage
    578
  • Abstract
    We report energy resolution, capacitance, and depletion voltage transients in neutron damaged coaxial Ge(HP) detectors fabricated from n-type and p-type germanium. The disordered region model for fast neutron degradation of germanium detectors was supported by 1) the large trapping cross sections observed (~ 10-11 cm2), 2) changes in the steady state charge on the traps over four decades of induced current density, and 3) the inability of the isolated defect hypothesis to self-consistently account for the ratio of thermal emission of holes in liquid nitrogen to liquid argon cooled detectors. The results of in-cryostat annealing of these detectors are reported.
  • Keywords
    Capacitance; Coaxial components; Current density; Detectors; Energy resolution; Germanium; Neutrons; Steady-state; Thermal degradation; Voltage;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1981.4331241
  • Filename
    4331241