DocumentCode
855532
Title
Electron Linear Accelerator Structures and Design for Radiation Therapy Machines
Author
Vaguine, Victor A.
Author_Institution
Varian Associates Inc., Palo Alto, California 94303
Volume
28
Issue
2
fYear
1981
fDate
4/1/1981 12:00:00 AM
Firstpage
1884
Lastpage
1888
Abstract
Requirements for straight- and bent-beam medical accelerator configurations are compared. Technical progress of straight-beam accelerators over the past 12 years is reviewed and the trade-off between optimized shunt impedance and maximum accelerating gradient in structure design is examined. Relative advantages of side-coupled and interlaced structures are delineated in the light of much higher permissible internal electric fields, recently demonstrated. A novel concept for a single-section standing-wave structure with multiple photon mode capability is presented.
Keywords
Acceleration; Assembly; Biomedical applications of radiation; Electron beams; Impedance; Linear accelerators; Magnetic heads; Magnetic separation; Manufacturing; X-rays;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1981.4331547
Filename
4331547
Link To Document