• DocumentCode
    856590
  • Title

    Ultrastable wavelength calibration reference constructed from an athermal Mach-Zehnder interferometer filter

  • Author

    Hao, J. ; Liaw, C.Y. ; Cai, Z. ; Gong, Y. ; Cheng, T.H. ; Lu, C.

  • Author_Institution
    Inst. for Infocomm Res.
  • Volume
    42
  • Issue
    24
  • fYear
    2006
  • Firstpage
    1422
  • Lastpage
    1424
  • Abstract
    A simple wavelength reference was constructed by illuminating an athermal silica-based Mach-Zehnder interferometer filter with a broadband light source. The nonlinear wavelength-voltage response of a Fabry-Perot tunable filter can be calibrated by using the constructed wavelength reference. The maximum wavelength deviation of the calibrated system with the Fabry-Perot tunable filter is 6times10-5 nm over a monitoring period of 12 h
  • Keywords
    Mach-Zehnder interferometers; calibration; measurement standards; optical fibre filters; 12 h; Fabry-Perot tunable filter; Mach-Zehnder interferometer filter; athermal Mach-Zehnder interferometer; broadband light source; nonlinear wavelength-voltage response; ultrastable wavelength calibration reference;