DocumentCode
864193
Title
Passive Neutron Multiplication Measurements
Author
Zolnay, A.S. ; Barnett, C.S. ; Spracklen, H.P.
Author_Institution
Lawrence Livermore National Laboratory P. O. Box 808 Livermore, CA 94550
Volume
30
Issue
1
fYear
1983
Firstpage
811
Lastpage
814
Abstract
We have developed an instrument to measure neutron multiplication by statistical analysis of the timing of neutrons emitted from fissionable material. This instrument is capable of repeated analysis of the same recorded data with selected algorithms, graphical displays showing statistical properties of the data, and preservation of raw data on disk for future comparisons. In our measurements we have made a comparison of the covariance to mean and Feynman variance to mean analysis algorithms to show that the covariance avoids a bias term and measures directly the effect due to the presence of neutron chains. A spherical assembly of enriched uranium shells and acrylic resin reflector/moderator components used for the measurements is described. Preliminary experimental results of the Feynman variance to mean measurements show the expected correlation with assembly multiplication.
Keywords
Algorithm design and analysis; Assembly; Detectors; Displays; Event detection; Instruments; Laboratories; Neutrons; Statistical analysis; Timing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1983.4332382
Filename
4332382
Link To Document