Title
A Stationary Positron Emission Ring Tomograph Using Bgo Detector and Analog Readout
Author
Burnham, C.A. ; Bradshaw, J. ; Kaufman, D. ; Chesler, D. ; Brownell, G.L.
Author_Institution
Massachusetts General Hospital, Physics Research Laboratory Boston, MA 02114
Volume
31
Issue
1
fYear
1984
Firstpage
632
Lastpage
636
Abstract
A single ring PET camera has been constructed to demonstrate the use of one dimensional scintillation camera techniques in positron tomographic design. A pseudo-continuous BGO detector is used. The detector is 46 cm ID, 52 cm OD, 2 cm wide and is made using 360 elements. A narrow light guide and 90 PM Tubes are used for the readout. The position of coincident events are found using unique scintillation camera logic.