• DocumentCode
    887469
  • Title

    Dose Enhancement Effects in a Shepherd Model 81-22 Cobalt-60 Irradiator

  • Author

    Zietlow, Thomas C.

  • Author_Institution
    Electronics Research Laboratory the Aerospace Corporation P. O. Box 92957 Los Angeles, California 90009
  • Volume
    34
  • Issue
    3
  • fYear
    1987
  • fDate
    6/1/1987 12:00:00 AM
  • Firstpage
    662
  • Lastpage
    666
  • Abstract
    The magnitude of the low energy component of the photon energy spectrum inside a J. L. Shepherd Model 81-22 Co-60 irradiator has been experimentally determined. Thermoluminescent dosimeters and a specialized ionization chamber were used to verify the usefulness of a lead (0.06") and aluminum (0.03") filter box to harden the Co-60 photon energy spectrum. The area of uniform dose as a function of distance from the Co-60 source inside the irradiator also has been determined. The results suggest two recommendations for use of the irradiator: the Pb/Al box should be used for all irradiated samples where low energy radiation can be a factor, and samples should be placed as close to the source as possible, while maintaining a uniform dose rate over the entire sample area.
  • Keywords
    Aerospace electronics; Aluminum; Electronic equipment testing; Geometry; Ionization chambers; Laboratories; Lead; Radiation effects; Semiconductor device testing; System testing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1987.4334698
  • Filename
    4334698