DocumentCode
887469
Title
Dose Enhancement Effects in a Shepherd Model 81-22 Cobalt-60 Irradiator
Author
Zietlow, Thomas C.
Author_Institution
Electronics Research Laboratory the Aerospace Corporation P. O. Box 92957 Los Angeles, California 90009
Volume
34
Issue
3
fYear
1987
fDate
6/1/1987 12:00:00 AM
Firstpage
662
Lastpage
666
Abstract
The magnitude of the low energy component of the photon energy spectrum inside a J. L. Shepherd Model 81-22 Co-60 irradiator has been experimentally determined. Thermoluminescent dosimeters and a specialized ionization chamber were used to verify the usefulness of a lead (0.06") and aluminum (0.03") filter box to harden the Co-60 photon energy spectrum. The area of uniform dose as a function of distance from the Co-60 source inside the irradiator also has been determined. The results suggest two recommendations for use of the irradiator: the Pb/Al box should be used for all irradiated samples where low energy radiation can be a factor, and samples should be placed as close to the source as possible, while maintaining a uniform dose rate over the entire sample area.
Keywords
Aerospace electronics; Aluminum; Electronic equipment testing; Geometry; Ionization chambers; Laboratories; Lead; Radiation effects; Semiconductor device testing; System testing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1987.4334698
Filename
4334698
Link To Document