• DocumentCode
    909179
  • Title

    Noise Distribution Due to Emission and Transmission Statistics in Positron Emission Tomography

  • Author

    Palmer, M.R. ; Bergstrom, M. ; Pate, B.D. ; Beddoes, M.P.

  • Author_Institution
    UBC/TRIUMF PET Program the University of British Columbia 2211 Wesbrook Mall Vancouver, B. C., V6T 2B5 Canada
  • Volume
    33
  • Issue
    1
  • fYear
    1986
  • Firstpage
    439
  • Lastpage
    442
  • Abstract
    Noise contributions to PET image data are due mainly to the limited counting statistics in the profiles of both emission and transmission scans. Statistical variance "images" were obtained with the UBC/TRIUMF PETT VI tomograph representing noise due to the corresponding emission and transmission scans. Phantom measurement variances were compared with those predicted analytically and found to be in close agreement. Variance contributions due to transmission noise are found to be most pronounced towards the centre of the image.
  • Keywords
    Analysis of variance; Attenuation measurement; Biomedical imaging; Councils; Imaging phantoms; Noise reduction; Pollution measurement; Positron emission tomography; Statistical distributions; Statistics;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1986.4337138
  • Filename
    4337138