DocumentCode
909203
Title
Figures of Merit for Different Detector (Configurations Utilized in High Resolution Positron Cameras
Author
Eriksson, L. ; Bergström, M. ; Bohm, Chr ; Holte, S. ; Kesselberg, M. ; Litton, J.
Author_Institution
Departments of Neuroradiology and Neurophysiology, Karolinska Hospital/Institute, Department of Physics, University of Stockholm, Sweden
Volume
33
Issue
1
fYear
1986
Firstpage
446
Lastpage
451
Abstract
A new positron camera system is currently being designed by our group. The goal is an instrument that can measure the whole brain with a spatial resolution of 5 mm FWHM in all directions. In addition to the high spatial resolution, the system must be able to handle count rates of 0.5 MHz or more in order to perform accurate fast dynamic function studies such as the determination of cerebral blood flow and cerebral oxygen consumption following a rapid bolus. An overall spatial resolution of 5 mm requires crystal dimensions of 6*6*L mm3, or less, L being the length of the crystal. Timing and energy requirements necessitate high performance photomultipliers. The identification of the small size scintillator crystals can currently only be handled in schemes based on the Anger technique, in the future possibly with photodiodes. In the present work different crystal identification schemes have been investigated. The investigations have involved efficiency measurements of different scintillators, line spread function studies and the evaluation of different coding schemes in order to identify small crystals.
Keywords
Blood flow; Cameras; Crystals; Detectors; Instruments; Photodiodes; Photomultipliers; Positrons; Spatial resolution; Timing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1986.4337140
Filename
4337140