• DocumentCode
    920753
  • Title

    Maximum likelihood positioning in the scintillation camera using depth of interaction

  • Author

    Gagnon, D. ; Pouliot, N. ; Laperrière, L. ; Therrien, M. ; Olivier, P.

  • Author_Institution
    Biomed. Eng., Montreal Heart Inst., Que., Canada
  • Volume
    12
  • Issue
    1
  • fYear
    1993
  • fDate
    3/1/1993 12:00:00 AM
  • Firstpage
    101
  • Lastpage
    107
  • Abstract
    Specific effects of the depth of interaction (DOI) on the photomultiplier (PM) response in an Auger gamma camera were quantified. The method was implemented and tested on a Monte Carlo simulator with special care to the noise modeling. Two models were developed, one considering only the geometric aspects of the camera and used for comparison, and one describing a more realistic camera environment. In a typical camera configuration and 140-keV photons, the DOI alone can account for a 6.4-mm discrepancy in position and 12% in energy between two scintillations. Variation of the DOI can still bring additional distortions when photons do not enter the crystal perpendicularly such as in slant hole, cone beam and other focusing collimators. With a 0.95-cm crystal and a 30° slant angle, the obliquity factor can be responsible for a 5.5-mm variation in the event position. Results indicate that both geometrical and stochastic effects of the DOI are definitely reducing the camera performances and should be included in the image formation process
  • Keywords
    biomedical equipment; cameras; radioisotope scanning and imaging; 0.95 cm; 140 keV; Auger gamma camera; Monte Carlo simulator; cone beam collimator; depth of interaction; event position; focusing collimators; maximum likelihood positioning; medical diagnostic imaging; noise modeling; nuclear medicine; obliquity factor; photomultiplier response; scintillation camera; slant hole collimator; stochastic effects; Cameras; Focusing; Monte Carlo methods; Optical collimators; Photomultipliers; Photonic crystals; Solid modeling; Stochastic processes; Testing; Working environment noise;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/42.222673
  • Filename
    222673