• DocumentCode
    928223
  • Title

    Performance Characterization of a Miniature, High Sensitivity Gamma Ray Camera

  • Author

    Olcott, P.D. ; Habte, Frezghie ; Foudray, Angela M. ; Levin, Craig S.

  • Author_Institution
    Molecular Imaging Program, Stanford Univ., Stanford, CA, USA
  • Volume
    54
  • Issue
    5
  • fYear
    2007
  • Firstpage
    1492
  • Lastpage
    1497
  • Abstract
    A compact, hand-held gamma camera with excellent intrinsic and extrinsic performance has been developed for the rapid identification and localization of sentinel lymph nodes during the surgical staging of cancer. The camera comprises a 5×5 cm2 field of view NaI (Tl) pixellated crystal array, a high sensitivity lead (2.0 cm thick) hexagonal parallel-hole collimator, a position sensitive photomultiplier tube (PSPMT), and a novel highly multiplexed electrical readout. The intrinsic energy resolution (12.1±2.0%) at 140 keV including edge crystals, extrinsic sensitivity (5 cps/μCi from 1-5 cm with a 24% energy window) and extrinsic spatial resolution (1.81±0.1 mm at 0.6 cm) facilitate rapid identification of a radioactive node. Using a node phantom we performed an ideal observer study to estimate the detectability of small spheres in the range of 3 mm to 8 mm with different exposure times. With a 5 seconds exposure the camera can detect a 3 mm diameter sphere at a depth of 3.6 cm containing 1 μCi of Tc-99m within a cold background, and a 4 mm diameter sphere containing 2 μCi at 2.6 cm depth within a warm background (>10:1 contrast) with a 5 second exposure duration.
  • Keywords
    biomedical equipment; cameras; gamma-ray apparatus; photomultipliers; position sensitive particle detectors; solid scintillation detectors; NaI(Tl) pixellated crystal array; cancer; hand-held gamma camera; hexagonal parallel-hole collimator; high sensitivity gamma ray camera; highly multiplexed electrical readout; position sensitive photomultiplier tube; sentinel lymph nodes; Cameras; Cancer; Collimators; Crystals; Energy resolution; Imaging phantoms; Lymph nodes; Oncological surgery; Photomultipliers; Spatial resolution; Camera technology; Monte Carlo simulation; NMIS; PET instrumentation; SPECT instrumentation; detectors; mammotomography; miniature radiation cameras; x-ray CT instrumentation;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2007.902367
  • Filename