• DocumentCode
    936162
  • Title

    Comments on "A cone-beam filtered backprojection reconstruction algorithm for cardiac single photon emission computed tomography" by G. T. Gullberg and G. L. Zeng

  • Author

    Wang, G. ; Lin, T.H.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., State Univ. of New York, Buffalo, NY, USA
  • Volume
    12
  • Issue
    2
  • fYear
    1993
  • fDate
    6/1/1993 12:00:00 AM
  • Firstpage
    383
  • Lastpage
    384
  • Abstract
    In the above-titled work by G. T. Gullberg and G. L. Zeng (ibid., vol.11, no.1, p.91-101, 1992), a fan-beam reconstruction formula of a noncircular scanning locus was derived and extended for half-scan cone-beam reconstruction. However, the reconstruction formula is not exact mathematically unless a necessary condition is satisfied. In this correspondence, the commenters derive this necessary condition and provide a geometrical explanation of the condition.<>
  • Keywords
    cardiology; computerised tomography; image reconstruction; radioisotope scanning and imaging; cardiac single photon emission computed tomography; cone-beam filtered backprojection reconstruction algorithm; fan-beam reconstruction formula; medical diagnostic imaging; noncircular scanning locus; nuclear medicine; Biomedical imaging; Diagnostic radiography; Hardware; Image quality; Image reconstruction; Jacobian matrices; Optical computing; Reconstruction algorithms; Single photon emission computed tomography; Supercomputers;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/42.232270
  • Filename
    232270